Smart A Endo Motor Woodpecker
Aliquota IVA applicata: 22%
BB mode and LB mode
BB Mode Developed
The file moves with continuous rotation at a 90° angle with a 1s interval between each movement. BB is short for "Block Bypass"; sometimes there are dentin debris or pulp remnants that block the apex, and this rotation with up and down movement can bypass the blockage.
LB Mode
The file moves with a gradual change of angle, moving in a circle of "set angle -2°, -1°, -0°, +1°" . LB is short for "ledge bypass". There are many reasons for ledge formation. It is necessary to bypass the ledge to locate the original root canal. First, ensure the file bypasses the ledge. Once the ledge has been bypassed, restart the switch to adjust the continuous movement state.
Impressive reciprocating mode
Adjustable rotation in reciprocating mode: adjustment range 5°-400°, adjustment interval 5°